Typical Atrial Flutter - When Do You Say You Have Got It



Typical Atrial Flutter - When Do You Say You Have Got It

Quick View
Credits:

Michael Peyrol, MD and Pascal Sbragia, MD .
Department of Cardiology, Centre Hospitalier Universitaire Nord, Marseille, France


Corresponding Author:  Michael Peyrol, MD, Department of Cardiology, Centre Hospitalier Nord, Chemin des Bourrelys, 13915 Marseille cedex 20, France.

Introduction

Right Atrial Flutter (AFL) is a common form of macro reentrant arrhythmia.1-3 In absence of previ- ous cardiac surgery, the reentry circuit is usually bounded anteriorly by the tricuspid annulus (TA) and posteriorly by the ostia of vena cava and Eu- stachian ridge. In this case, AFL is consensually called "typical" AFL and is highly dependent of the cavotricuspid isthmus (CTI)2. The CTI is a criti- cal channel which represents the predominant area of slow conduction of the circuit. Therefore, this narrow isthmus has become the universally ac- cepted target for ablation of typical AFL.2-4 If abla- tion is carried out during AFL the first "intuitive" procedure endpoint is arrhythmia termination. Although this latter was initially thought to be an acceptable endpoint for ablation procedure, bidi- rectional CTI conduction block is actually considered as the gold standard endpoint for elimination of typical AFL recurrence.5,6 Indeed, Schumacher et al. found a 9% recurrence rate after bidirec- tional CTI block achievement, 54% recurrence rate after unidirectional CTI block and 100% recur- rence rate when persistent slow conduction across CTI was noted after RF application on the CTI.7

As cardiac electrophysiology laboratories use a wide variation of catheters, sheaths, mapping systems or energy sources for typical AFL ablation, techniques to validate bidirectional CTI conduction block are also varied.8-10 In fact, for the last 15 years, many elec- trophysiological criteria were proposed at this end: reversal in activation of right atria around the abla- tion line, extent of prolongation of the transisthmus interval after ablation, documentation of a corridor of double potentials (DPs) along the ablation line, differential atrial pacing, analysis of the local uni- polar and bipolar electrograms, incremental atrial pacing and changes in paced P wave morphology. We review these electrophysiological criteria commonly used for CTI conduction block validation.


Reversal in the Right Atrial Activation around the Ablation Line

Complete bidirectional CTI block across the line of ablation was first identified with 20-pole Halo- catheter recordings during coronary sinus (CS) ostium/low lateral right atria (LLRA) pacing.5,11 Halo-catheter was placed around the TA for CTI conduction mapping (fig 1A). In patients in si- nus rhythm before RF application, conduction across the CTI was confirmed by demonstrating bidirectional RA depolarization wavefronts with collision in the RA free wall or interatrial septum during CS ostium/LLRA pacing respectively (fig 1B). Complete bidirectional conduction block was determined by a reversal in the RA activation around the TA. In this case, clockwise only RA ac- tivation around the TA when pacing lateral to the line of ablation and counter clockwise only RA ac- tivation when pacing septal to the line of ablation were observed (fig. 1C).5,11 Low septal atrial pac- ing is sometimes difficult to perform due to high pacing thresholds. In this setting, right ventricular pacing and resulting retrograde atrial septal and CTI activation can be useful to validate conduction block.12 Otherwise, pacing at slow rates (i.e ≤ 600 ms) is mandated because CTI conduction block may be functional and rate-dependant in some patients.

Figure 1A: Radioscopic view (60° LAO) of standard catheter position on the high, mid, and low lateral RA wall, mid sep- tum in the His bundle region, lower septum, and proximal coronary sinus


Main pitfall of this maneuver was that, in some patients, rapid conduction across the crista termi- nalis would result in a modification of the activa- tion sequence around the Halo-catheter when not covering the CTI .10 In this case, transverse con- duction through the crista terminalis, septal to lat- eral, during CS pacing would depolarize the low lateral RA and thus anticipate the RA depolarization sequence on distal Halo-catheter mimicking incomplete CTI block while this latter is present. Scaglione et al. found trans-crista terminalis con- duction in 3/12 patients (25%)13 On account of this phenomenon Anselme et al. reported that in 7 of 38 patients (18%), the RA activation sequence on Halo-catheter was not sufficient to detect com- plete CTI block.14Then, difficulties in accurately positioning Halo-catheter were reported to occur in up to 10% of the cases.14 This pitfall might be associated with significant increase in procedure duration and fluoroscopy exposure. Finally, RF applications may slow conduction velocity at the CTI without blocking it. Consequently, the two wave fronts might collide on the very LLRA (Low Lateral Right Atrium)dur- ing CS pacing mimicking complete CTI block on Halo-catheter.

Figure 1B: Left, Pacing from the low lateral RA. Mid lateral RA is activated first, then high lateral RA. Activation of the septum occurs through dual wave fronts: Low region of sep- tum is activated caudocranially and upper region is activated craniocaudally. Right, pacing from the PCS. Activation of the septum is caudocranial. Activation of the lateral RA is mainly caudocranial, low lateral RA being activated before mid and high lateral RA. Nevertheless, high lateral RA is activated be- fore mid lateral RA, owing to a craniocaudal wave front


Figure 1C: Pacing from the low lateral RA. After ablation, all the septum is activated craniocaudally via a single clockwise front. Right, pacing from the PCS. The lateral RA is activated craniocaudally via a single counterclockwise front.Complete CTI conduction block is confirmed. H, M, and L Lat= high, mid, and low lateral RA wall; H, M1, M2 Sept L Sept = high, mid septum 1 and 2 and low septum; PCS = proximal coro- nary sinus.Reproduced from Cauchemez B et al. Circulation 1996;93:284-294


Extent of Prolongation of the Transisthmus Interval after Ablation

Extent of prolongation of transisthmus conduc- tion intervals during atrial pacing was studied to predict complete CTI conduction block after creation of a line of ablation.15 Transisthmus conduction intervals were measured between the stimulus artifact and the local atrial activation recorded from the proximal CS electrode during pacing from LLRA in counterclockwise direction and between stimulus artifact and the local atrial activation recorded from the pair of electrodes positioned lateral to the ablation line during pac- ing from CS ostium in clockwise direction. This maneuver was performed before and after RF application. In 57 consecutive patients, Oral etal. reported that an increase ≥ 50% in transithmus intervals during LLRA and CS pacing predicted complete bidirectional CTI block. Conversely, complete CTI block was not observed in any pa- tient if transisthmus conduction interval did not increase by at least 50 %. Diagnostic accuracy and positive predictive value of this technique for identification of complete CTI block was 92% and 90 % respectively. On account of these data, au- thors supported this criterion only as an additional indicator for evaluation of CTI ablation.


Corridor of Double Potentials along the Line of Ablation

In experimental studies, double potential (DP) sep- arated with an isoelectric interval were shown to validate local conduction block under the exploring bipolar electrode.16 During CS/LLRA pacing, a corridor of parallels DPs separated with an iso- electric interval along the CTI ablation line, from the RV edge to the inferior vena cava edge strongly support the presence of bidirectional CTI conduc- tion block.17-20 Specific characteristics of DPs pre- dicting CTI block were analyzed. On the one hand, Tada et al demonstrated that DP interval, mea- sured from the peak of the first component of the DP to the peak of the second component, reflected a local gap/incomplete CTI block when measured.<90 ms whereas DP interval >110 ms was associated with complete CTI block (fig 2).19 A separation >110 ms between the two components of the DP along the whole line of ablation was found to have a pre- dictive positive value of 100 % for complete CTI block. Others studies confirmed these data.17,18,20 On the other hand, when DP interval was between 90 and 110 ms, a maximal variation ≤15 ms in the DP intervals, an isoelectric segment separating the two components of the DP and a negative polar- ity of the second component of the DP were found to be additional reliable indicators of complete CTI block.19 Otherwise, during typical AFL or CS os- tium pacing, residual gaps could be identified by local electrograms with a single or a fractionated potential centered on, or spanning, the isoelectric interval of adjacent DP.8,17 Main pitfall of this tech- nique is that the demonstration of the existence of a corridor of large DPs at the level of the line of block needs to map strictly the whole ablation line and is frequently difficult to prove. Indeed, Anselme et al. found that in 39 % of patients, low amplitude signals, ambiguous or atypical DPs due to multiple RF applications or due to the complex anatomy of the isthmus were recorded.14In these cases, accuracy of this technique was significantly reduced.

Figure 2: Increase in degree of splitting within DPs during coronary sinus pacing before (left and middle) and after (right) complete CTI conduction block. The recording abla- tion catheter was positioned at exactly the same site in all three panels. Note in the middle panel a DP interval = 96 ms without an isoelectric interval between the two components of DP and incomplete CTI block on Halo-catheter. In left pan- el, DP interval increased to 126 ms with an isoelectric interval and complete CTI block on halo-catheter.(E9 through E1 = electrograms recorded on Halo catheter; CS = coronary sinus; Abl = ablation catheter) Reproduced from Tada et al. JACC 2001;38:750 –5


Differential Pacing

As mentioned above, interpretation of DP on CTI after multiple RF applications is frequently difficult. Otherwise, separation of the two components is variable not always reaching the 110 ms cut-off.19 In these cases, differential pacing is helpful to differentiate CTI block from persistent slow conduction.21-24 The four pacing sites, two septal and two lateral to the line of ablation, and their relation with anatomical structures are shown (fig 3A).

Figure 3A: Schematic representation (LAO 45° projection) of the four pacing sites. Reproduced from Chen et al. Circula- tion. 1999;100:2507-2513


This dynamic pacing maneuver was used to determine whether the two components of the DP are due to a single depolarization wavefront penetrating through the ablation line or to two opposite wavefronts recorded on either sides of a complete line of block. If the two components of DP are induced by persistent slow conduction through the CTI, then pacing farther away from the ablation line results in an increase in the in- terval between stimulus and the two components of DP in similar manner (fig 3B).

Figure 3B: Validation of complete bidirectional CTI block us- ing differential pacing. Note the greater activation delay at site D when pacing at site A than when pacing at site B (AD delay>BD delay), and greater activation delay at site A when pacing at site D than when pacing at site C (DA delay>CA delay)


Therefore the in- terval between the two electrograms remains unchanged. Conversely, if the two components of DP are induced by two opposite wavefronts around the ablation line with complete conduction block, then preceding maneuver results in similar effect on electrogram corresponding to the same side of the ablation line but that due to activation on other side of the line is advanced. Therefore the separation between the two electrograms decreases. In a large series of 255 consecutive patients, long-term results of CTI ablation with conduction block validated with this simplified method were comparable to those obtained with standard approaches.24 Otherwise, this technique, using only one quadripolar mapping electrode in addition to the ablation catheter, rules out the use of a mul- tipolar catheter and thus reduces the cost of the procedure. Differential pacing can be performed using a multipolar Halo catheter (2-8-2 spacing) and the ablation catheter as recently reported by Pastor et al..25 The Halo catheter is coiled in the RA, as a reference electrode, and the ablation catheter, as the recording electrode, is positioned at either the posterior or anterior edge of the CTI (fig 3C).

Figure 3C: LAO view of the two electrode catheters, 24 pole Halo catheter and ablation catheter, set-up for CTI conduction assessment as reported by Pastoretal. Differential pacing can be performed using A3 and A4 electrode pairs or S3 and S4 electrode pairs with the ablation catheter, as the recording electrode, positioned at the posterior or anterior edge of the CTI respectively. Reproduced from Pastoretal. Europace. 2010; 12:1290-5


The authors considered that CCW CTI block was achieved when, during pacing with the distal elec- trodes A4 of the Halo-catheter: - septal (posterior) RA activation was descending, - the late compo- nent of the CTI EGM (recorded with the ablation catheter) was later than the low septal EGM (re- corded with the Halo-catheter), and pacing at a distance from the CTI (A4 vs A3 in fig 3C) short- ened the interval between the pacing stimulus and the late CTI EGM. Conversely, CW was considered to be achieved when: - lateral (anterior) RA acti- vation was completely descending when pacing with the ablation catheter at the septal end of the CTI, - criteria of differential pacing at the RA sep- tum (S4 vs S3 in fig 3C), with recording electrode at the anterior end of the CTI, was validated. Pastoretal. reported a low rate of AFL recurrence (3%) during a mean follow-up of 37 months when bidirectional CTI block was validated with the aforementioned criteria.25 With this approach, CTI block can both be validated with differential pacing and activation sequence of the RA around the ablation line.


Unipolar and Bipolar Local Electrograms

Unipolar electrograms are commonly used for the guidance of ablation of arrhythmia such as focal ventricular/atrial tachycardia or accessory pathways.26,27 Background for the use of unipolar electrograms is that RS morphology of the unipo- lar electrogram is associated with propagation of the wave front through the exploring electrode, whereas positive monophasic R unipolar electro- gram is characteristic of the termination of the activation wavefront.28,29 Villacastin et al. studied unipolar atrial electrograms recorded along the line of ablation in 45 consecutive patients with typical AFL.30The authors found that local unipo- lar atrial electrograms obtained at the LLRA dur- ing CS pacing changed from RS, rS, or QS before CTI ablation to R or Rs after clockwise CTI block was obtained (fig 4A). Similar changes were observed when counterclockwise conduction block was obtained. Main pitfall of this technique was that the mapping catheter has to be placed very accurately, in the very close vicinity of the abla- tion line, increasing both time of procedure and radiation exposure. Bipolar atrial electrograms morphology were also evaluated as an addi- tional criterion for complete CTI block.19,32 Com- plete CTI could be validated when a change in the morphology of bipolar electrograms adjacent and lateral to the line of block from an RS pattern to a QR pattern was observed. A 100% correlation was found between bipolar atrial electrograms morphology and previously validated change of the atrial activation sequence and differential pacing.32This report was concordant with study from Tada et al who showed that reversal of bi- polar electrogram polarity recorded lateral to the line of ablation was a simple and accurate indicator of CTI conduction block.19

Figure 4: Local electrograms registered during CS ostium pac- ing with ablation catheter used as exploring electrode record- ing bipolar (Bip) and unipolar (Unip) signals at the very close lateral side of ablation line before and after RFA. We observe a change from rS pattern (pre-RFA) to a monophasic R wave (post-RFA) at Unip consistent with CTI conduction block. Note post-RFA double potentials observed at Bip and loss of caudo-cranial activation of the lateral wall on Halo-catheter. (MS = mid interatrialseptum; HS = high interatrial septum; Roof = roof of right atrium; HAL = high anterolateral; MAL = mid anterolateral; LAL = low anterolateral.) Reproduced from Villacastin J et al. Circulation.2000 Dec 19;102(25):3080-5


Incremental Pacing

Incremental pacing from CS or LLRA is the most recently local electrogram-based maneuver pro- posed for validation of CTI conduction block.33 It consists in monitoring of DP distance along the CTI line during CS and LLRA pacing from 600ms to 250 ms. Authors hypothesized that a significant increase in DP distance during incremental pacing would indicate persistent slow conduction through the isthmus, while continuity in DP distance would validate complete CTI conduction block. Measurements of DP distance and response to this pacing maneuver should be performed at a CTI point with a maximum separation of the two potentials. In this observational study, a <20 ms increase in DP distance during incremental pacing maneuver confirmed CTI block (fig 5). Conversely, a >20 ms increase in DPs distance indicated persistence of conduction through the isthmus and necessity for further RF applications. Main advantage of this maneuver is that extensive mapping of the whole ablation line is not mandated. Limitation of this technique was reached in presence of transversal conduction across the crista terminalis which anticipated the second DP component and might lead to misdiagnose complete CTI block. In this setting, incremental pacing should be combined to assessment of polarity of second DP component unipolar electrogram obtained during LLRA/CS ostium pacing.

Figure 5: Incremental pacing from CS ostium with cycle lengths at 600 ms, 500 ms, 400 ms, 300 ms and 250 ms. Top panel: Significant > 20 ms increase in DPs distance during incremental pacing maneuver consistent with persistent slow conduction across the ablation line. Bottom panel: Absence of increase > 20 ms in DP distance with incremental pacing indicating CTI conduction block. (CTI = cavotricuspid isthmus; DPs = double potentials; Halo = 20-pole Halo catheter; MAP = ablation catheter; CS = coronary sinus.) Reproduced from Bazan V et al. J Cardiovasc Electrophysiol. 2010 Jan;21(1):33-9


Monitoring of changes in the paced P wave morphology

Changes in the paced P wave morphology have been reported as a marker of CTI block achievement.34,35 Shahetal. reported that, in a prospective group of 30 patients, CTI block achievement was associated with a terminal positivity of the P wave during LLRA pacing, best seen in inferior leads, in 29 patients.35 Otherwise the authors reported that conduction recovery through the isthmus, validated with intracardiac mapping, could be recognized by the regression of this morphologic P wave change.35 However, this technique provides information only on counter clockwise conduction recovery.


Cavotricuspid isthmus ablation in the era of electroanatomical mapping

Three-dimensional electroanatomical mapping systems are currently used in the majority of cardiac electrophysiology laboratories for the guidance of various arrhythmia ablations such as atrial fibrillation, non CTI-dependant atrial flutter or ventricular tachycardia. Electroanatomical mapping technologies were demonstrated to be safe and effective to guide placement of linear lesion at the level of the CTI.36-39 All authors reported significant reduction of radiation exposure with 3-D mapping systems compared to conventional approach.36-40 Visualization of CTI conduction on activation maps during LLRA/CS ostium pacing, in association with registration of wide > 100 ms DPs along the line of ablation, might help the vali- dation of CTI block.36,40 Main limitation with the non-fluoroscopic approach was that the material cost per procedure is significantly increased (42%) compared to the conventional strategy with the Halo catheter.37 Otherwise, no favorable impact on procedure duration, number of RF applica- tions or clinical outcome was encountered with 3-D mapping system which may explain that this technology remains a marginal tool for typical AFL ablation in absence of previous complex cardiac surgery.36-39


Observation period after achievement of bidirectional conduction block

Recovery of CTI conduction can occur after initial achievement of complete CTI block .9,21,23,41 Most of recurrences of CTI conduction occur within 20-30 minutes. Bruetal.found recovery of conduction across the isthmus in 5/35 patients (14%) within 20 minutes after initial successful CTI ablation.41 Shah et al. observed transient CTI conduction block in 20/40 patients (50%) during observation period.23In 39 patients, Chen et al. re- ported resumption of conduction once in 10 patients (25.6%) and twice in 3 patients (7.7%) within 30 minutes after end of initial CTI ablation.21 These reports support the necessity of a 30-minutes ob- servation period after achievement of CTI block to detect early recovery of conduction. Isoproterenol administration was proposed to unmask transient CTI block.42 However, no evidence was found this was superior to the 30-minutes observation period.9

Conclusions

Complete CTI conduction block provides an objective endpoint for typical AFL ablation. All electrophysiological criteria proposed to validate this procedural endpoint have their own limitations. Therefore, in order to reduce arrhythmia recurrence, validation of CTI conduction block should combine different techniques. In our experience, reversal in the RA activation sequence during CS/ LLRA pacing, validated by differential pacing, actually represents the most accurate and commonly used criteria. In difficult cases, interpretation of local unipolar or bipolar electrograms adjacent to the line of ablation may be helpful to validate complete CTI conduction block.

Disclosures

No disclosures relevant to this article were made by the authors.

References

1. Granada J, Uribe W, Chyou PH et al. Incidence and predictors of atrial flutter in the general population. J Am Coll Cardiol. 2000; 36:2242-6.
2. Saoudi N, Cosío F, Waldo A,et al. A classification of atrial flut- ter and regular atrial tachycardia according to electrophysiologi- cal mechanisms and anatomical bases; a Statement from a Joint Expert Group from The Working Group of Arrhythmias of the European Society of Cardiology and the North American Society of Pacing and Electrophysiology. Working Group of Arrhythmias of the European of Cardiology and the North American Society of Pacing and Electrophysiology. Eur Heart J. 2001; 22:1162-82.
3. Da Costa A, Thévenin J, Roche F et al. Results from the Loire- Ardéche-DrÔme-Isére-Puy-de-Dôme (LADIP) trial on atrial flut- ter, a multicentric prospective randomized study comparing amiodarone and radiofrequency ablation after the first episode of symptomatic atrial flutter. Circulation.2006; 114:1676-81.
4. Natale A, Newby KH, Pisanó et al. Prospective randomized comparison of antiarrhythmic therapy versus first-line radiofre- quency ablation in patients with atrial flutter. J Am Coll Cardiol. 2000; 35:1898-904.
5. Nakagawa H, Lazzara R, Khastgir T et al. Role of the tricus- pid annulus and the eustachian valve/ridge on atrial flutter: Relevance to catheter ablation of the septal isthmus and a new technique for rapid identification of ablation success.Circula- tion.1996; 94:407-24.
6. Schwartzman D, Callans DJ, Gottlieb CD et al. Conduction block in the inferior vena caval-tricuspid valve isthmus: associa- tion with outcome of radiofrequency ablation of type I atrial flut- ter. J Am Coll Cardiol. 1996; 28:1519-31.
7. Schumacher B, Pfeiffer D, Tebbenjohanns J et al. Acute and long-term effects of consecutive radiofrequency applications on conduction properties of the subeustachian isthmus in type I atri- al flutter. J Cardiovasc Electrophysiol.1998; 9:152-63.
8. Takahashi A, Shah DC, Jaïs P et al. How to ablate typical atrial flutter. Europace. 1999; 1:151-5.
9. Cosío FG, Awamleh P, Pastor A et al. Determining inferior vena cava-tricuspid isthmus block after typical atrial flutter abla- tion. Heart Rhythm.2005; 2:328-32.
10. O'Neill MD, Jais P, Jönsson A, et al. An approach to catheter ablation of cavotricuspid isthmus dependent atrial flutter. Indian Pacing Electrophysiol J. 2006; 6:100-10.
11. Cauchemez B, Haissaguerre M, Fischer B et al. Electrophysi- ological effects of catheter ablation of inferior vena cava-tricus- pid annulus isthmus in common atrial flutter. Circulation. 1996; 93:284-94.
12. Vijayaraman P, Kok LC, Wood MA et al. Right ventricular pacing to assess transisthmus conduction in patients undergo- ing isthmus-dependent atrial flutter ablation: a new useful tech- nique? Heart Rhythm. 2006; 3:268-72.
13. Scaglione M, Riccardo R, Calo L et al. Typical atrial flutter ablation: conduction across the posterior region of the inferior vena cava orifice may mimic unidirectional isthmus block. J Cardiovasc Electrophysiol 2000; 11:387­95.
14. Anselme F, Savouré A, Cribier A et al. Catheter ablation of typical atrial flutter: a randomized comparison of 2 methods for determining complete bidirectional isthmus block. Circulation.2001; 103:1434-9.
15. Oral H, Sticherling C, Tada H et al. Role of transisthmus conduction intervals in predicting bidirectional block after ablation of typical atrial flutter. J Cardiovasc Electrophysiol.2001; 12:169- 74.
16. Allessie MA, Bonke FIM. Schopman FJC: Circus movement in rabbit atrial muscle as a mechanism of tachycardia. Circ Res 1973;33:54-62.
17. Shah DC, Haïssaguerre M, Jaïs P et al. Simplified electrophys- iologically directed catheter ablation of recurrent common atrial flutter. Circulation. 1997; 96:2505-8.
18. Tai CT, Haque A, Lin YK et al. Double potential interval and transisthmus conduction time for prediction of cavotricuspid isthmus block after ablation of typical atrial flutter. J Interv Card Electrophysiol.2002; 7:77-82.
19. Tada H, Oral H, Sticherling C et al. Double potentials along the ablation line as a guide to radiofrequency ablation of typical atrial flutter. J Am Coll Cardiol. 2001; 38:750-5.
20. Poty H, Saoudi N, Nair M et al. Radiofrequency catheter ablation of atrial flutter. Further insights into the various types of isthmus block: application to ablation during sinus rhythm. Circulation.1996; 94:3204-13.
21. Chen J, de Chillou C, Basiouny T et al. Cavotricuspid isthmus mapping to assess bidirectional block during common atrial flut- ter radiofrequency ablation. Circulation. 1999; 100:2507-13.
22. Shah D, Haïssaguerre M, Takahashi A et al. Differential pac- ing for distinguishing block from persistent conduction through an ablation line. Circulation. 2000; 102:1517-22.
23. Shah DC, Takahashi A, Jaïs P et al. Local electrogram-based criteria of cavotricuspid isthmus block. J Cardiovasc Electro- physiol. 1999; 10:662-9.
24. Maury P, Raczka F, Gaty D et al. Radio-frequency ablation of atrial flutter: long-term results and predictive value of cavo- tricuspid isthmus bidirectional block as determined by a simpli- fied technique. Cardiology. 2008; 110:17-28.
25. Pastor A, Núñez A, Guzzo G, de Diego C, Cosío FG. A simple pacing method to diagnose postero-anterior (clockwise) cavo- tricuspid isthmus block after radiofrequency ablation. Europace. 2010; 12:1290-5.
26. Haïssaguerre M, Dartigues JF, Warin JF et al. Electrogram patterns predictive of successful catheter ablation of accessory pathways. Value of unipolar recording mode. Circulation. 1991; 84:188-202.
27. Delacretaz E, Soejima K, Gottipaty VK et al. Single catheter determination of local electrogram prematurity using simultane- ous unipolar and bipolar recordings to replace the surface ECG as a timing reference. Pacing Clin Electrophysiol. 2001; 24:441-9.
28. Spach MS, Miller WT III, Geselowitz DB, et al. The discontinu- ous nature of propagation in normal canine cardiac muscle: evi- dence for recurrent discontinuities of intracellular resistance that affect the membrane currents. Circ Res. 1981;48:39­54.
29. Spach MS, Miller WT, Miller-Jones E, et al. Extracellular po- tentials related to intracellular action potentials during impulse conduction in anisotropic canine cardiac muscle. Circ Res. 1979; 45:188­204
30. Villacastin J, Almendral J, Arenal A et al. Usefulness of uni- polar electrograms to detect isthmus block after radiofrequency ablation of typical atrial flutter. Circulation. 2000; 102:3080-5.
31. Lin YJ, Tai CT, Huang JL et al. Characteristics of virtual uni- polar electrograms for detecting isthmus block during radiofre- quency ablation of typical atrial flutter. J Am Coll Cardiol. 2004; 43:2300-4.
32. Andronache M, de Chillou C, Miljoen H et al. Correlation be- tween electrogram morphology and standard criteria to validate bidirectional cavotricuspid block in common atrial flutter abla- tion. Europace. 2003; 5:335-41.
33. Bazan V, Martí-Almor J, Perez-Rodon J et al. Incremental pac- ing for the diagnosis of complete cavotricuspid isthmus block during radiofrequency ablation of atrial flutter. J Cardiovasc Electrophysiol. 2010; 21:33-9.
34. Hamdam MH, Kalman JM, Barron HV et al. P wave morphol- ogy during right atrial pacing before and after atrial flutter abla- tion—a new marker for success. Am J Cardiol 1997;79:1417­20.
35. Shah DC, Takahashi A, Jaïs P et al. Tracking dynamic conduction recovery across the cavotricuspid isthmus. J Am Coll Cardiol. 2000 ;35:1478-84.
36. Willems S, Weiss C, Ventura R et al. Catheter ablation of atrial flutter guided by electroanatomic mapping (CARTO): a randomized comparison to the conventional approach. J Cardiovasc Electrophysiol. 2000; 11:1223-30.
37. Hindricks G, Willems S, Kautzner J et al. Effect of electroanatomically guided versus conventional catheter ablation of typical atrial flutter on the fluoroscopy time and resource use: a prospective randomized multicenter study. J Cardiovasc Electrophysiol. 2009; 20:734-40.
38. Schneider MA, Ndrepepa G, Dobran I et al. LocaLisa catheter navigation reduces fluoroscopy time and dosage in ablation of atrial flutter: a prospective randomized study. J Cardiovasc Electrophysiol. 2003;14:587-90.
39. Kottkamp H, Hügl B, Krauss B et al. Electromagnetic versus fluoroscopic mapping of the inferior isthmus for ablation of typical atrial flutter: A prospective randomized study. Circulation. 2000;102:2082-6.
40. Schneider MA, Ndrepepa G, Zrenner B et al. Noncontact mapping-guided ablation of atrial flutter and enhanced-density mapping of the inferior vena caval-tricuspid annulus isthmus. Pacing Clin Electrophysiol. 2001;24:1755-64.
41. Bru P, Duplantier C, Bourrat M et al. Resumption of right atrial isthmus conduction following atrial flutter radiofrequen- cy ablation. Pacing Clin Electrophysiol. 2000; 23:1908-10.
42. Nabar A, Rodriguez LM, Timmermans C et al. Isoproterenol to evaluate resumption of conduction after right atrial isthmus ablation in type I atrial flutter. Circulation. 1999; 99:3286-91.


Biosense Webster
event date
Introduction to AFib
Ablation Specialist

View Ablation Specialists