Percutaneous Interventional Electrophysiology and Real-Time Three-Dimensional Transoesophageal Echocardiography in the Cardiac Electrophysiology Laboratory: Is it an Exceptional Technique During Electrophysiological Study Especially at Atrial Fibrillation?




Percutaneous Interventional Electrophysiology and Real-time Three-Dimensional Transoesophageal Echocardiography in the Cardiac Electrophysiology Laboratory: Is it an Exceptional Technique During Electrophysiological Study Especially at Atrial Fibrillation?
¹Mustafa Yildiz, ²Cuneyt Toprak
1Department of Cardiology, Istanbul University Cardiology Institue, Istanbul, 2Department of Cardiology, Kartal Koşuyolu Yuksek Ihtisas Educational and Research Hospital, Istanbul.

Abstract
Percutaneous electrophysiological interventions such as atrial fibrillation and ventricular tachycardia ablation are usually monitored by fluoroscopy and electroanatomical mapping systems. Electroanatomical mapping systems lacks soft tissue contrast and adequate visualization of the target area such as right ventricular out-flow tract and left atrium. Recently, real-time 3D transoesophageal echocardiography (RT-3D TEE) has emerged as an important method for visualizing cardiac structures such as left atrium, left atrial appendix, interatrial septum, pulmonary veins, Marshall ligament and mitral valve annulus during invasive procedures. This review aims to describe the RT-3D TEE for the guidance of percutaneous interventional electrophysiological study especially at atrial fibrillation in the cardiac electrophysiology laboratory.

Corresponding Address : Dr. Mustafa Yildiz, M.D., Ph.D, Prof., Cardiologist, Internal Medicine Specialist and Physiologist, Bayar Cad, Gulbahar Sok, A Blok, D6, Kozyatağı/Istanbul.

Introduction
Percutaneous electrophysiological interventions including catheter ablation of supraventricular and/or ventricular tachycardia are usually monitored by fluoroscopy and electroanatomical mapping systems.1,2 Use of the electroanatomical mapping systems including Ensite NaVx and/or CARTO has been shown to reduce fluoroscopic time and radiation dose. Although this systems can use for complex procedures such as atrial fibrillation and ventricular tachycardia, it lacks soft tissue contrast and adequate visualization of the target area such as right ventricular out-flow tract and left atrium.2 Two Dimensional (2D) intracardiac echocardiography can also use for the guidance of electrophysiological interventions especially ablation of the atrial fibrillation. This technique improves the efficacy of electrophysiological procedures by identifying anatomical structures and integrating this information with electrophysiological parametersand/or three dimensional (3D) reconstructions of computed tomography/magnetic resonance imaging data.3 This method is limited by its invasive nature, using large sheaths, 11 F sheath for a 10 F intracardiac echocardiography catheter or 9 F sheath for an 8 F intracardiac echocardiography catheter,3 with possible bleeding complications. 2 Recently, in line with improvement in the quality of echocardiographic imaging, real-time 3D transoesophageal echocardiography (RT-3D TEE), is not off-line 3D reconstruction, is available to facilitate electrophysiological procedures such as transseptal puncture without fluoroscopy.2 It is possible to obtain cross-sectional visualization of the left atrium, left atrial appendix, pulmonary veins such as left superior pulmonary vein and its anatomical variations, mitral valve annulus, thrombus, and prosthetic valves less invasively with RT-3D TEE 2 (Figure 1,Figure 2), which are not possible with 2D TEE.
Figure 1. It is possible to obtain excellent visualization of the anatomical area with RT-3D TEE during interventional electrophysiology [LA: Left atrium, RA: Right atrium, LAA: Left atrial appendix, LSPV: Left superior pulmonary vein, LSPVa (anatomical variation): Left superior pulmonary vein a, LSPVb (anatomical variation): Left superior pulmonary vein b, MV: Mitral valve, ML: Marshall ligament, IAS: Interatrial septum, THROM: Thrombus, PMV: Prosthetic mitral valve, PAV: Prosthetic aortic valve, PTV: Prosthetic tricuspide valve]



Figure 2. Visualization of the left atrial appendix, its thrombus and transseptal puncture (MV: Mitral valve, LA: Left atrium, LAA: Left atrial appendix, THROM: Thrombus, PMV: Prosthetic mitral valve, PAV: Prosthetic aortic valve, IAS: Interatrial septum)



RT-3D TEE was performed in the supine position using an iE33 echocardiography system (Philips Medical Systems, Andover, USA) with a RT-3D Matrix transducer (X7-2t).4,5 Initially, gain settings were optimized using the narrow angle gain mode without the need for electrocardiography allowed us to obtain RT-3D images of an approximately 30°X60° pyramidale volume. To magnify the pyramidal volume, 3D zoom mode was used for anatomical and pathological tissue visualization such as mitral valve, pulmonary veins and thrombus. Image pages were divided automatically by the instrument into 2 mm or 5 mm squares for measurement of image size. The size of the images including pulmonary veins, left atrial appendix, mitral valve, patent foramen ovale and thrombus were estimated by these squares.5 Although RT-3D TEE allows for unique perspectives for catheter-based procedures such as atrial fibrillation, some disadvantages of RT-3D TEE must be considered as well. One disadvantage is prolonged TEE probe exposure time most likely due to the learning curve because of the new technique. Secondly, this technique is still limited bya relatively low temporal and spatial resolution when compared with conventional 2D TEE. Additionally, it need for off-line reconstruction and analysis for 3D distances and volumes. Third, the anaesthesiology personnel are required during the procedure of atrial fibrillation is performed for sedation. Finally, the high cost of equipment such as the transducer and the 3D-capable echo system may be limiting factors for a number of echo laboratories.
Conclusions
Implementation of RT-3D TEE in percutaneous electrophysiological interventions especially atrial fibrillation, left atrial appendage closure and transseptal puncture results in reduction of fluoroscopic time and radiation dose, may reduce complication ratio, and may improve in-hospital and long term outcomes.
Disclosures
No disclosures relevant to this article were made by the authors.
References
  • Bhakta Deepak , Miller John M .Principles of electroanatomic mapping. Indian Pacing Electrophysiol J . 2008; 8 (1):32–50. [ PubMed]
  • Balzer Jan , Kelm Malte , Kühl Harald P .Real-time three-dimensional transoesophageal echocardiography for guidance of non-coronary interventions in the catheter laboratory. Eur J Echocardiogr . 2009; 10 (3):341–9. [ PubMed]
  • Biermann Jürgen , Bode Christoph , Asbach Stefan .Intracardiac Echocardiography during Catheter-Based Ablation of Atrial Fibrillation. Cardiol Res Pract . 2012; 2012 ():–. [ PubMed]
  • Azran Marc S , Romig Christopher B , Locke Andrew , Whitley William S .Echo rounds: application of real-time 3-dimensional transesophageal echocardiography in the percutaneous closure of a mitral paravalvular leak. Anesth. Analg. . 2010; 110 (6):1581–3. [ PubMed]
  • Yildiz Mustafa , Duran Nilüfer Ekşi , Gökdeniz Tayyar , Kaya Hasan , Ozkan Mehmet .The value of real-time three-dimensional transesophageal echocardiography in the assessment of paravalvular leak origin following prosthetic mitral valve replacement. Turk Kardiyol Dern Ars . 2009; 37 (6):371–7. [ PubMed]


Biosense Webster
event date
Introduction to AFib
Ablation Specialist

View Ablation Specialists